GNR AreX D desktop retained austenite analyzer with the lid open
GNR Analytical Instruments Group

Desktop Retained Austenite Analyzer — AreX D by GNR

The AreX D is a benchtop retained-austenite analyzer built to ASTM E 975-13. It measures down to 1% by volume within minutes, weighs 37 kg and is air-cooled, so no cooling water is needed.

Retained austenite

AreX D — technical detail

AreX D — key specifications

  • Benchtop XRD for measuring retained austenite per ASTM E 975-13
  • Down to 1 vol% in a few minutes · typical measuring time 180 seconds
  • Mo-anode X-ray tube, 50 W, air-cooled, with Zr filter
  • 2θ range 21.5°–44.5° · DECTRIS MYTHEN2 R 1D microstrip detector
  • Video camera for sample alignment · sample holder 200 × 150 × 70 mm
  • Safety: door interlock, thick steel frame and lead shielding, radiation below 1 mSv/year

AreX D — size and installation

490 × 550 × 360 mm, 37 kg, 110/220 V 50/60 Hz. The X-ray tube is air-cooled, so no cooling water is needed.

Used for heat-treatment quality control and gear/bearing parts — GNR’s example measures three points on a gear tooth (pitch diameter, top land, top corner).

AreX D precision (120-second measurement)

Retained austenite (%)Standard deviation (S.D.)
300.8
150.6
100.4
30.2

Talk to Two Theta about your project

Request a quoteCall